DATOSPIR MICRO User’s Manual
208
Chapter 8: Upkeep, Preventative and Corrective Maintenance
511-A00-MU2 • REV. 2.06
Like any equipment, particularly if for medical applications,
the DATOSPIR MICRO spirometer requires upkeep and
maintenance aimed firstly at the safety of patients, operators
and the environment, and secondly, at ensuring the reliability
and precision of the functions for which it has been developed.
All this leads to a series of routines that must be completed.
8.1 UPKEEP
Upkeep is aimed at ensuring the correct working order of the
equipment. The person undertaking it requires no special technical
skills except knowledge of the functions and handling of the
equipment.
The equipment user himself must normally perform this. The
operations to be completed are as follows:
CLEANING THE TURBINE TRANSDUCER
Given that the turbine is the part exposed directly to the patient,
it must be kept in perfect physical and hygienic conditions. To do
so, proceed as follows:
1
The turbine is removed from the equipment housing by
pressing slightly so that is comes away from its fixtures.
2
Wash the turbine with water and soap, avoiding solvents or
abrasive substances that may damage the components. Given
that its reliability depends on the condition of the turbine, make
sure it is not damaged.
3
Once rinsed, avoid salt deposits by rinsing again with distilled
water.
4
You can then leave it to dry at room temperature.
5
Reassemble the turbine in the housing.
Summary of Contents for DATOSPIR MICRO A
Page 1: ...DATOSPIR MICRO 511 A00 MU2 REV 2 06 2019 03 USER MANUAL SPIROMETER...
Page 6: ...Index DATOSPIR MICRO User s Manual 6 511 A00 MU2 REV 2 06...
Page 8: ...DATOSPIR MICRO User s Manual 8 Declaration of Conformity 511 A00 MU2 REV 2 06...
Page 14: ...Safety DATOSPIR MICRO User s Manual 14 511 A00 MU2 REV 2 06...
Page 64: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 64 511 A00 MU2 REV 2 06 U...
Page 67: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 67 511 A00 MU2 REV 2 06 U...
Page 127: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 127 511 A00 MU2 REV 2 06 U...
Page 131: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 131 511 A00 MU2 REV 2 06 U...
Page 133: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 133 511 A00 MU2 REV 2 06 U...
Page 137: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 137 511 A00 MU2 REV 2 06 U...
Page 192: ...Chapter 5 Technical Specifications DATOSPIR MICRO User s Manual 192 511 A00 MU2 REV 2 06...
Page 198: ...Chapter 6 Operating Principles DATOSPIR MICRO User s Manual 198 511 A00 MU2 REV 2 06...
Page 218: ...Appendix 1 Electromagnetic Compatibility DATOSPIR MICRO User s Manual 218 511 A00 MU2 REV 2 06...