Chapter 7: Spirometry Technique
DATOSPIR MICRO User’s Manual
202
511-A00-MU2 • REV. 2.06
although at least one of the elements of the biological signal, the
volume or the flow, must be reproduced together or separately.
Along these lines, the syringes holding several litres provide a
suitable signal and the flow generators can be used to assess
the precision and errors in flow measurement. Among the
most suitable calibration appliances is the so-called explosive
decompressor, which consists of a 4 or 5-litre pressurised, one-
atmosphere chamber fitted with a fast opening for the sudden
expulsion of an identical volume to that of the chamber. This
simulates the forced expiration of a person and, with suitable
resistances at different levels of obstruction fitted to the output
tube, the signal is similar to that of a patient with slight, moderate
or severe obstruction in the air flow. Hence, the volume and
flow measurement can be examined. If this is not possible, the
working order of the appliance must be checked using «control
individuals». i.e. people close to the laboratory and whose
cooperation is attainable, to carry out correct spirometry with
easy and scarce variability (Table I), so that their spirometry can
be regularly reproduced and compared with the previous data.
Hence, errors can be detected that would have to be significant
in size, as the variability of spirometry itself prevents small
differences in the measuring of volume and flow from being
detected (see Table I).”
“Under normal working conditions, calibration using the volume
signal provided by a manual syringe will be performed on a daily
basis. The signal provided by the syringe must be produced
with different pulses to check whether the flow reading remains
steady, as the appliance must always integrate the signal in the
same volume - that provided by the syringe signal - whatever
the suddenness of the injection manoeuvre, with the upper limit
of the flow range precisely measured (close to the real value) by
the instrument in question not being exceeded. Calibration with
a dynamic signal produced by the explosive decompressor or the
measuring of spirometry on control individuals, as mentioned
above, may be performed relatively less often. It is wise to
perform calibration using the decompressor every fortnight in the
case of pneumotachometers. Given that spirometry with control
Summary of Contents for DATOSPIR MICRO A
Page 1: ...DATOSPIR MICRO 511 A00 MU2 REV 2 06 2019 03 USER MANUAL SPIROMETER...
Page 6: ...Index DATOSPIR MICRO User s Manual 6 511 A00 MU2 REV 2 06...
Page 8: ...DATOSPIR MICRO User s Manual 8 Declaration of Conformity 511 A00 MU2 REV 2 06...
Page 14: ...Safety DATOSPIR MICRO User s Manual 14 511 A00 MU2 REV 2 06...
Page 64: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 64 511 A00 MU2 REV 2 06 U...
Page 67: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 67 511 A00 MU2 REV 2 06 U...
Page 127: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 127 511 A00 MU2 REV 2 06 U...
Page 131: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 131 511 A00 MU2 REV 2 06 U...
Page 133: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 133 511 A00 MU2 REV 2 06 U...
Page 137: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 137 511 A00 MU2 REV 2 06 U...
Page 192: ...Chapter 5 Technical Specifications DATOSPIR MICRO User s Manual 192 511 A00 MU2 REV 2 06...
Page 198: ...Chapter 6 Operating Principles DATOSPIR MICRO User s Manual 198 511 A00 MU2 REV 2 06...
Page 218: ...Appendix 1 Electromagnetic Compatibility DATOSPIR MICRO User s Manual 218 511 A00 MU2 REV 2 06...