Chapter 7: Spirometry Technique
DATOSPIR MICRO User’s Manual
204
511-A00-MU2 • REV. 2.06
“Multi-centre study in Barcelona
Variable Sex Equation (Age 6-20) R
SEE
FVC
M 0.0.0.05728E-3.21 0.947 0.443
F
0.0.0.03550E-3.04 0.935
0.313
FEV1
M 0.0.0.07148E-2.91 0.945
0.378
F
0.0.0.02955E-2.87 0.940
0.263
*FEV1/FVC% M 0.593E - 0.113P + 81.60
F
0.026T + 82.60
FEF 25-75% M 0.038T + 0.140E - 4.33
0.832 0.796
F
0.046T + 0.051E - 4.30 0.789 0.651
PEF
M 0.075T + 0.275E - 9.08 0.907 1.073
F
0.073T + 0.134E - 7.57
0.879 0.831
FEF 50%FVC
(*)
M
0.017T + 0.157E + 0.029P - 2.17
0.856 0.811
F
0.046T + 0.067E - 4.17
0.803 0.669
FEF 75%FVC
(*)
M 0.024T + 0.066E - 2.61
0.760 0.562
F 0.027T + 0.032E - 2.68
0.709 0.507
Variable Sex Equation (Age 20-70)
R SEE
FVC
M
0.0678T - 0.0147E - 6.05
0.72
0.530
F
0.0454T - 0.0211E - 2.83
0.75
0.403
FEV1
M
0.0499T - 0.0211E - 3.84
0.75
0.444
F
0.0317T - 0.0250E - 1.23
0.82
0.307
*FEV1/FVC % M
- 0.1902E + 85.58
F
- 0.224E - 0.1126P + 94.88
FEF 25-75% M
0.0392T - 0.0430E - 1.16
0.55
1.000
F
0.0230T - 0.0456E + 1.11
0.70
0.680
PEF
M
0.0945T - 0.0209E - 5.77
0.47
1.470
F
0.0448T - 0.0304E + 0.35
0.47
1.040
FEF 50%FVC
(*)
M
0.0517T - 0.0397E - 2.40
0.47
1.300
F
0.0242T - 0.0418E + 1.62
0.56
0.925
FEF 25%FVC
(*)
M
0.0190T - 0.0356E - 0.14
0.63
0.620
F
0.02T-0.031E-0.0062P-0.21
0.76 0.405
*FEV1/PEF M
6.64
F
7.77
*FEV1/FEV0.5 M
1.45
F
1.50
*FEF50/FIF50
(*)
M
0.66
F
0.88
*PEF/PIF M
1.39
F
1.42
*FEV1/FIV1 M
0.80
F
0.89
Summary of Contents for DATOSPIR MICRO A
Page 1: ...DATOSPIR MICRO 511 A00 MU2 REV 2 06 2019 03 USER MANUAL SPIROMETER...
Page 6: ...Index DATOSPIR MICRO User s Manual 6 511 A00 MU2 REV 2 06...
Page 8: ...DATOSPIR MICRO User s Manual 8 Declaration of Conformity 511 A00 MU2 REV 2 06...
Page 14: ...Safety DATOSPIR MICRO User s Manual 14 511 A00 MU2 REV 2 06...
Page 64: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 64 511 A00 MU2 REV 2 06 U...
Page 67: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 67 511 A00 MU2 REV 2 06 U...
Page 127: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 127 511 A00 MU2 REV 2 06 U...
Page 131: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 131 511 A00 MU2 REV 2 06 U...
Page 133: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 133 511 A00 MU2 REV 2 06 U...
Page 137: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 137 511 A00 MU2 REV 2 06 U...
Page 192: ...Chapter 5 Technical Specifications DATOSPIR MICRO User s Manual 192 511 A00 MU2 REV 2 06...
Page 198: ...Chapter 6 Operating Principles DATOSPIR MICRO User s Manual 198 511 A00 MU2 REV 2 06...
Page 218: ...Appendix 1 Electromagnetic Compatibility DATOSPIR MICRO User s Manual 218 511 A00 MU2 REV 2 06...