Appendix 1: Electromagnetic Compatibility
DATOSPIR MICRO User’s Manual
215
511-A00-MU2 • REV. 2.06
Guidance and manufacturer’s declaration – electromagnetic emissions
DATOSPIR MICRO
is intended for use in the electromagnetic environment specified below.
The costumer or the user of
DATOSPIR MICRO
should assure that it is used in such an
environment.
Tests EN-IEC 61000-4-4 y -4-5 are applicable to AC/DC power inputs, and input/output
signal. Test EN-IEC 61000-4-6 additionally applies to patient connections. Test EN-IEC
61000-4-11 only applies to AC power inputs.
Immunity test
EN-IEC 60601 test level
Compliance level
Electromagnetic
environment –
Guidance
Electrical fast
transient/burst
EN-IEC 61000-4-4
±2 kV for power supply
lines and earth
±1 kV for input/output
lines
Frecuency 100 kHz
Not applicable
Not applicable
DATOSPIR MICRO
works with
batteries.
The input/output
line cables are
shorter than
3 meters long.
Surge
EN-IEC 61000-4-5
±0.5, ±1 kV differential
mode (Line to line).
±0.5, ±1, ±2 kV common
mode (Line to ground).
Not applicable
Not applicable
DATOSPIR MICRO
works with
batteries.
Conducted RF
EN-IEC 61000-4-6
3 Vrms
de 150KHz a 80 MHz
6 Vrms in ISM bands
80% AM a 1kHz
3 Vrms
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
EN-IEC 61000-
4-11
0% Ut ; 0.5 cycles: 0º,
45º, 90º, 135º, 180º,
225º, 270º y 315º.
0% Ut; 1 cycle
70% Ut; 25/30 cycles: 0º
0% Ut ; 250/300 cycles
(5 segundos)
Not applicable
DATOSPIR MICRO
works with
batteries.
NOTE Ut is the a.c. mains voltage prior to application of the test level.
Summary of Contents for DATOSPIR MICRO A
Page 1: ...DATOSPIR MICRO 511 A00 MU2 REV 2 06 2019 03 USER MANUAL SPIROMETER...
Page 6: ...Index DATOSPIR MICRO User s Manual 6 511 A00 MU2 REV 2 06...
Page 8: ...DATOSPIR MICRO User s Manual 8 Declaration of Conformity 511 A00 MU2 REV 2 06...
Page 14: ...Safety DATOSPIR MICRO User s Manual 14 511 A00 MU2 REV 2 06...
Page 64: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 64 511 A00 MU2 REV 2 06 U...
Page 67: ...Chapter 2 Model A Operating DATOSPIR MICRO User s Manual 67 511 A00 MU2 REV 2 06 U...
Page 127: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 127 511 A00 MU2 REV 2 06 U...
Page 131: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 131 511 A00 MU2 REV 2 06 U...
Page 133: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 133 511 A00 MU2 REV 2 06 U...
Page 137: ...Chapter 3 Models B and C Operation DATOSPIR MICRO User s Manual 137 511 A00 MU2 REV 2 06 U...
Page 192: ...Chapter 5 Technical Specifications DATOSPIR MICRO User s Manual 192 511 A00 MU2 REV 2 06...
Page 198: ...Chapter 6 Operating Principles DATOSPIR MICRO User s Manual 198 511 A00 MU2 REV 2 06...
Page 218: ...Appendix 1 Electromagnetic Compatibility DATOSPIR MICRO User s Manual 218 511 A00 MU2 REV 2 06...