16-2 Replacing Fuses
1
Turn the power off and disconnect the power cord from the unit.
2
The fuse holder is located at the rear of the unit. Insert a slot screwdriver or similar
tool from the top of the fuse holder and pull the fuse holder towards you.
3
Remove the fuses from the fuse holder and check if any are burned out.
4
If there are no burned-out fuses, return the fuse holder to its original position. Re-
place burned-out fuse(s) with spare fuse(s) (8A).
5
Return the fuse holder to its original position in the main unit.
6
Insert the power cord back into the main unit.
*
Should the fuse(s) blow out repeatedly after replacement, contact your Shimadzu
representative.