23
5) Specifying settings for STEP (Stepped drying) mode
In stepped drying mode it is possible to specify separate drying conditions for up to a maximum of 5 steps.
Specifying a separate drying temperature and measuring time for each step makes it possible to determine
the optimum drying conditions for any sample type.
In Step 1, timed ending mode must be specified as the ending condition. In Step 2 and subsequent steps, it
is possible to select between either automatic ending mode or timed ending mode.
*
If automatic ending mode is selected in Step 2 or later, that step becomes the final step. When timed ending
mode is selected and a value of ‘0’ is specified as the measuring time, the previous timed ending step becomes
the final step.
Display
Key(s) used
Operating instructions
With the display of the weight in grams, press the
CONDITION
key.
If a password has been specified, the message ‘PASS’ will be
displayed to indicate that the password should be entered. Use
the
key,
key and
ENTER
key to enter the password. Follow
the instructions under step
7
in “9-2-7 Specifying a password” to
set or delete a password.
The
‘CONDITIONS’
display area will begin to flash. Press the
SELECT
key repeatedly until
‘MODE’
begins to flash.
With the
‘MODE’
display flashing, press the
ENTER
key. This will
cause control to shift to measuring mode selection and for the
currently specified measuring mode to begin flashing.
Press the
SELECT
key repeatedly until ‘STP’ begins to flash.
*
If ‘STP’ is already flashing, then proceed directly on to the next
step.
When ‘STP’ flashes, press the
ENTER
key. This will cause ‘STP’ to
change from a flashing to a constantly lit display. Control will shift
to the specification of the measuring time for Step 1, and the
measuring time will flash.
Press the
key or
key to select the desired measuring time.
The measuring time may be specified in 1-minute increments in a
range anywhere from 1 to 240 minutes. Note that here a time of
20 minutes has been specified for Step 1.
When the desired measuring time flashes, press the
ENTER
key. The
measuring time will change from a flashing to a constantly lit
display, control will shift to the specification of the drying
temperature, and the drying temperature will begin to flash.
Press the
key or
key to select the desired drying
temperature. The drying temperature may be specified in 1-
degree increments in a range from 30°C to 200°C (30°C to 180°C
in 'TT' mode). Note that here a temperature of 105°C has been
specified for Step 1.
When the desired drying temperature flashes, press the
ENTER
key.
The step displayed will change from ‘STP 1’ to ‘STP 2’ and the
units area for timed ending mode (‘min’) or automatic ending
mode (‘%’) will begin to flash.
Step
1
2
3
4
5
6
7
8
9
0
CONDITION
ENTER
ENTER
SELECT
SELECT
ENTER
ENTER
SALES & TECHNICAL SUPPORT
QAQC LAB www.qclabequipment.com TEL (804) 318-3686
Summary of Contents for MOC-120H
Page 2: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Page 10: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Page 75: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Page 76: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...