46
Display
12-2-3 Preforming predictive measurements
Operating instructions
Set the mode to predictive measuring mode. Refer to 8.2.2 “7 Specifying
settings for PREDICT (Predictive measuring) mode” for instructions.
Specify the same settings for the predicted value convergence range and
drying temperature as those used in preparatory measuring mode (subsections
12.2.1 and 12.2.2).
Example:
Predicted value convergence range:
0.5%
Drying temperature:
120°C
Set the bias obtained through the procedures described in 12.2.1 and 12.2.2
(shown here as 0.60%). (See “8-2-4 Bias”for setting.)
Place the sample onto the sample pan and begin measurement. (In the
example shown here, the initial weight is 10.213 grams.)
Completion of measurement
The predicted measurement will then be displayed.
Step
1
2
3
4
5
. . . . .
SALES & TECHNICAL SUPPORT