![Shimadzu MOC-120H Instruction Manual Download Page 36](http://html1.mh-extra.com/html/shimadzu/moc-120h/moc-120h_instruction-manual_1256928036.webp)
24
Display
Key(s) used
Step
A
B
C
D
E
F
G
H
ENTER
ENTER
ENTER
SELECT
CONDITION
SELECT
ENTER
Operating instructions
Pressing the
SELECT
key will cause the units to switch between
minutes (‘min’) and a percentage (‘%’). Press the
SELECT
key until the
desired unit is displayed.
When the desired unit for the ending
condition begins to flash,
press the
ENTER
key. This will cause control to shift to the
specification of the ending
condition of Step 2 and the numerical
field will flash.
*
Note that this step becomes the final step if automatic ending
mode (i.e., a percentage) is specified here.
Press the
key or
key to select the desired ending
conditions. Note that here the timed ending at 20 minutes has
been specified. (Also note that if automatic ending mode has
been selected, a value from 0.01 to 0.1 percent may be specified
in 0.01-percent increments.
*
Note that the previous step becomes the final step if timed
ending mode is selected and a measuring time of ‘0’ is specified
here.
*
Note that continuous drying (i.e., a drying time of 12 hours)
cannot be specified in this mode. The maximum drying time for
each step is 240 minutes.
When the desired ending
condition
begin to flash, press the
ENTER
key. This will cause the display of the ending
condition
to change
from a flashing to a constantly lit display, for control to shift to the
specification of the drying temperature for Step 2, and the drying
temperature will begin to flash.
Press the
key or
key to select the desired drying
temperature. Note that here a temperature of 100°C has been
specified.
When the desired drying temperature is displayed, press the
ENTER
key. The step displayed will change from ‘STP 2’ to ‘STP 3’ and
control will shift to the specification of ending condition. Go back
to step
A
and repeat all subsequent steps up to this step until
settings have been specified for all the drying steps.
When the settings for the final drying step have been specified,
‘MODE’
will begin to flash.
If you wish to specify other measuring
conditions
settings, press
the
SELECT
key to select the menu.
To exit from specifying measuring
conditions
settings, press the
CONDITION
key instead.
. . . . .
SALES & TECHNICAL SUPPORT
QAQC LAB www.qclabequipment.com TEL (804) 318-3686
Summary of Contents for MOC-120H
Page 2: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Page 10: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Page 75: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Page 76: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...