25
6) Specifying settings for COMPARE (Preparatory measuring) mode
Preparatory measuring mode is used to calculate the bias (i.e., the difference between the measurement
results which would be obtained in Standard drying & Automatic ending mode and predictive measuring
mode) needed for predictive measurements.
Prior to predictive measurements, this mode should be used first to seek the bias. (For further information,
see “
Preparatory measuring mode” and “
Predictive measuring mode”.)
Display
Key(s) used
Operating instructions
With the display of the weight in grams, press the
CONDITION
key.
If a password has been specified, the message ‘PASS’ will be
displayed to indicate that the password should be entered. Use
the
key,
key and
ENTER
key to enter the password. Follow
the instructions under step
7
in “9-2-7 Specifying a password” to
set or delete a password.
The
‘CONDITIONS’
display area will begin to flash. Press the
SELECT
key repeatedly until
‘MODE’
begins to flash.
With the
‘MODE’
display flashing, press the
ENTER
key. This will
cause control to shift to measuring mode selection and for the
currently specified measuring mode to begin flashing.
Press the
SELECT
key repeatedly until ‘CMP’ begins to flash.
*
If ‘CMP’ is already flashing, then proceed directly on to the next
step.
When ‘CMP’ flashes, press the
ENTER
key. ‘CMP’ will change from a
flashing to a constantly lit display and you will next be required to
specify the predicted value convergence range settings (see “
Predicted value convergence range”(section 12)). The current
setting for the predicted value convergence range will begin to flash.
Press the
key or
key to specify the desired value. Any
value of 0.1 to 9.9 percent in increments of 0.1 percent may be
specified. Note that here a value of 0.5% has been selected.
When the desired value is displayed, press the
ENTER
key. The
display of the predicted value convergence range will change from
a flashing to a constantly lit display, and control will shift to the
automatic ending settings, and the setting specifying the degree of
change in moisture content over a 30-second period will begin to flash.
Press the
key or
key to select the desired automatic
ending condition. Automatic ending condition may be specified in
0.01-percent increments anywhere in a range from 0.01 to 0.1
percent. Note that here a value of 0.05% has been selected.
When the desired automatic ending condition has been selected,
press the
ENTER
key. The specified setting will change from a
flashing to a constantly lit display, and
‘MODE’
will begin to flash.
If you wish to specify other measuring conditions settings, press
the
SELECT
key to select the menu.
To exit from specifying measuring conditions settings, press the
CONDITION
key instead.
Step
1
2
3
4
5
6
7
8
9
0
A
CONDITION
ENTER
ENTER
SELECT
SELECT
ENTER
ENTER
SELECT
CONDITION
Note: Read Section 12 before use of this part.
SALES & TECHNICAL SUPPORT
QAQC LAB www.qclabequipment.com TEL (804) 318-3686
Summary of Contents for MOC-120H
Page 2: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Page 10: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Page 75: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Page 76: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...