
FlexPS Product Guide
Page 67
3
Calibration & setup
Introduction
FlexPS calibration is done, using Senstar’s Universal Configuration Module (UCM). The UCM is a
window-based software application that performs the calibration, setup, maintenance and
diagnostic functions for Senstar’s line of intrusion detection sensors. The UCM connects directly to
the FlexPS processor via USB. Network based processors can also connect remotely via the
Network Manager.
Senstar recommends that the initial calibration be done at the processor location using a direct
USB connection to the UCM. An enclosure tamper condition must exist to enable UCM
communication via a USB connection.
FlexPS calibration is a four stage process that requires:
•
Adjusting the processor’s Gain Settings
•
Adjusting the processor’s Cut Alarm Settings
•
Adjusting the processor’s Climb Alarm Settings
•
Testing the installation
Begin by using the Calibrate tool to observe the sensor’s response over the full length of the zone.
Once the sensor’s response is consistent along the length of the zone, you setup the processor’s
Cut Alarm Settings and Climb Alarm Settings. Finally, test the system to ensure that the detection
meets the site requirements.
Note
Consult the online help for detailed information on UCM operation.
Note
For network based systems, after completing the initial calibration
Senstar recommends running a magnitude response plot through the
Network Manager during the first periods of inclement weather. If the
weather causes an unacceptable number of nuisance alarms, you
can increase the Threshold setting to exclude the effects of the
weather.
For standalone systems that encounter an unacceptable number of
nuisance alarms during inclement weather, increase the Threshold
setting to exclude the effects of the weather.
After increasing the Threshold, retest the system to ensure that the
sensor’s detection meets your security requirements.