63 |
P a g e
R e v 5 - 1 9
1.
Take a Thickness Verification Unit (TVU) (
See Image at Right
) and place the probe head inside
the TVU until it stops. Insert a battery into the location as
shown in the image – this will turn on a rotating piece of film
inside the TVU.
2.
Turn on the
SpecMetrix®
software.
3.
Select the appropriate NIST Recipe for the TVU.
TVU
4.
Run the NIST recipe to make sure the values being read are within an acceptable tolerance
range to the NIST traceable value shown on the TVU.