43 |
P a g e
R e v 5 - 1 9
Lane Graph
If there is a [Lane Definition] provided, then this will represent the material from edge to edge (as
discovered by reflectance for SA Traversing or Photo Eyes used by PLC Traversing). If no [Lane
Definition] is provided, then this will represent the total width defined across the machine, and the
material evaluation will be a subset of the default number of lanes. Every individual light reading taken
and evaluated while the probe is in each “Lane” will be averaged together to yield a single lane
average. Each change of direction the traverse head makes will reset the ‘current traverse’ calculation
as the probe moves across the web. If the probe is stopped in a lane, the data will be recorded as a
single value regardless of how long the probe is motionless.
Die Graph
This represents the total width of the machine/roller, which is divided into a predetermined number of
regions/lanes/dies at the time of installation. Every individual light reading taken and evaluated while
the probe is in each “Lane” will be averaged together to yield a single lane average. Each change of
direction the traverse head makes will reset the ‘current traverse’ calculation as the probe moves across