Laser Detection Setup
Make sure laser is set to factory settings. (Hold the down
button and press the up button 5 times you should see an
“RST” on the LCD of the sensor, press up to get a “no” on
the LCD, then press “up” again and you will see “yes”, then
press “down” this will get you back to factory settings).
Move the sensor off of film, press the side button for 1 sec.
(It will flash “set”). Then press and hold side button till
“set” goes away. Do this for both sensors.
Network and Machine Communications
The SpecMetrix® Display Unit motherboards are designed with 2 or 3 Ethernet ports for various network
& peripheral communication needs. The Ethernet ports are configured for communication by Sensory
Analytics technicians or by plant IT engineers. When multiple devices need to communicate across
different networks or subnets, it is imperative to make sure the correct Ethernet cables are plugged into
the appropriate Ethernet port.
Plant network engineers are permitted to change the login configuration to use plant network
authentication. The requirement for any user account that will be given access to the SpecMetrix®
Display Unit is to have local admin access as well as proper access to read/write to any network path
that is defined as a file repository. It is permitted, and recommended, to use file mapping to network
paths.
For the main electrical cable, run a properly terminated (Line, Neutral, and Ground) power cable to the
side of the ICU to the clamshell connector provided on the panel mount. (connections are labeled). All
internal power is already wired appropriately.
Emergency Stop
SpecMetrix
Traversing systems come with an Emergency Stop (E-
Stop) push button (refer to provided electrical schematics for
wiring).
When the E-Stop is engaged, it will disengage controls to the
servo drives causing them to cease motion. A signal is also sent
to ‘Stop’ the Inline software from measuring.
While the E-Stop is engaged the servo drives will not be able to
move the probe head, and SA Inline cannot measure. The probe
head can be moved by hand when the E-Stop is engaged.
Ensure the probe is free from obstruction before releasing the E-
Stop. The E-Stop is released by turning the red mushroom head to
the “right” which will release the button and pop it out. Once the E-Stop is released, the probes will
move to their park position.