60 |
P a g e
R e v 5 - 1 9
System Maintenance
Suggested Cleaning Interval
The cleaning intervals are based upon general factory condition. If the box is placed in areas where
significant dust or particle density is high, or should you notice performance or temperature issues and
you may need to shorten the suggested interval cleaning cycle.
Environment
Maintenance Interval *
Clean Room / Office / Lab environment
Quarterly
Factory Environment
Monthly
Harsh Factory Environment
Weekly
*
- These are recommended minimum intervals. You may choose to do more frequently based on your
environment.
Maintenance Procedures for Lab and In-line systems
We suggest using compressed canned air for dust/particle removal. Never use water or liquids to clean
electronic components.
*** WARNING: Make sure power is turned off to electronics before performing any maintenance. ***
1)
Clean fan ports with compressed air (blow outward).
2)
Check particle filter, replace if necessary.
3)
Using Compressed air, blow off all circuit boards and electronics. Make sure dust is evacuated
from the box.
4)
Visually check wires for damage/fray and bad or loose connections.
Lamp Replacement for In-line Systems
There is one illumination lamp in the system for each optical probe. These lamps can be found in the
aluminum blocks identified by the yellow circles in the following picture (Figure 62).