Specifications
Model: Sensor Electronics Corporation
SEC MILLENIUM Infrared Hydrocarbon Gas Detector
Available gases:
Propane
Propylene
Methane
n-Butane
Ethane
Gasoline
Ethanol
Isopropyl Alcohol
Ethylene
Methanol
Aeromatic 150
Pentane
Hexane
Cyclopentane
Isobutane
Toluene
Methyl Amyl Ketone
P-Xylene
MEK
Tert-Butyl Acetate
Please note that this list is not all-inclusive. The SEC
MILLENIUM can be calibrated for most hydrocarbons,
provided a calibration gas is available. For more
information please contact Sensor Electronics Corporation.
Part Numbers:
Methane PN: 49000000100L12
(0-100% LEL)
Methane PN: 49000000050L12
(0-50% LEL)
Methane PN: 49000000100V12
(0-100% VOL)
Propane PN: 49000100100L12
(0-100% LEL)
Propane PN: 49000100100V12
(0-100% VOL)
Propane PN: 49000100100U12
(0-100% UEL)
Aeromatic 150 PN: 49000200100L12
(0-100% LEL)
Ethane PN: 49000300100L12
(0-100% LEL)
Ethanol PN: 49000400100L12
(0-100% LEL)
Ethylene PN: 49000500100L12
(0-100% LEL)
Gasoline PN: 49000600100L12
(0-100% LEL)
Hexane PN: 49000700100L12
(0-100% LEL)
Isobutane PN: 49000800100L12
(0-100% LEL)
Isopropyl Alcohol (IPA) PN: 49000900010L12
(1-100% LEL)
Methanol PN: 49001000100L12
(0-100% LEL)
N-Butane PN: 49001100100L12
(0-100% LEL)
Pentane PN: 49001200100L12
(0-100% LEL)
Methyl Amyl Ketone PN: 49001400100L12
(0-100% LEL)
Cyclopentane PN: 49002500100L12
(0-100% LEL)
Propylene PN: 49002900100L12
(0-100% LEL)
Toluene PN: 49003700100-L12
(0-100% LEL)
P-Xylene PN: 4900400100-L12
(0-100% LEL)
Tert-Butyl Acetate PN: 49003400100-L12 (0-100% LEL)
Detection Method: Diffusion - Optional sample draw
(requires a minimum of 1 liter per minute flow rate.)
Output (analog):
4-20 mA (Source type),
max. 1000 Ohm load at 24 VDC supply voltage
Response Time:
T50 < 5 seconds
T90 < 10 seconds
Construction:
316 stainless steel
SEC Millenium
Infrared Gas Detector
Document #1460001 Revision A
Sensor Electronics Corporation
12730 Creek View Avenue, Savage, MN 55378 U.S.A. • (800) 285-3651 • (952) 938-9486 • FAX: (952) 938-9617
Email: [email protected] • website: www.sensorelectronics.com
Sensor Electronics Corporation reserves the right to alter specifications without prior notice.
© 11/16
Unit Status Chart
Current Output
Status
4-20 mA
Normal measuring mode
0.0
mA
Unit Fault
0.2
mA
Reference channel fault
0.4
mA
Analytical channel fault
0.8
mA
Unit warm up
1.0
mA
Optics fault
1.2
mA
Zero drift fault
1.6
mA
Calibration fault
2.0
mA
Unit spanning
2.2
mA
Unit Zeroing
4.0
mA
Zero gas level
5.6
mA
10% LEL
8.0
mA
25% LEL
12
mA
50% LEL
16
mA
75% LEL
20
mA
100% LEL
20.1 – 23 mA Over range (>100%)
Accuracy:
+/- 3% LFL, 0 to 50% LFL (Lower Flammable Limit)
+/- 5% LFL, 51 to 100% LFL
Operating Temperature Rating:
-40˚ to +70˚C at 0 to 99% RH (non-condensing)
Operating Range:
18 to 32 VDC measured at the detector head
Power Consumption:
5 Watts Max
Max Current Draw: (at 24VDC)
Average: 210 mA Peak: 400 mA
Approvals:
C22.2 No. 152-M1984 (R1997)
Performance Tested
Class 1, Division 1, Groups B, C and D
Conforms to UL2075, Methane 0-100% LEL
Installation Category: Cat. I, Pollution Degree 2
Weight: 5 lbs. (2.3 kg.)
Summary of Contents for SEC 3000
Page 15: ...Doc 1460003 Rev A Operation Manual SEC 3000 Electro Chemical Gas Sensor Page 12 of 13 ...
Page 54: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 4 SEC 3100 SEC Millenium Mounting ...
Page 55: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 5 SEC 3100 SEC 3000 Mounting ...
Page 56: ...TB 1 J1 TB 2 J2 1 6 1 6 1 4 1 6 TB 5 J5 TB 6 J6 ...
Page 148: ...SEC 3120 Wiring TB 1 P3 TB 2 P6 1 6 1 6 1 2 3 4 1 2 3 4 7 6 5 4 3 2 1 TB 3 P1 TB 4 P5 ...
Page 156: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 399: ...SEC P N 1460014 Rev ECO A 000209 Page 11 of 26 ...
Page 400: ...SEC P N 1460014 Rev ECO A 000209 Page 12 of 26 ...
Page 431: ...Part Number 1460005 Revision C ...
Page 433: ...Part Number 1460005 Revision C ...
Page 434: ...Part Number 1460005 Revision C ...
Page 435: ...Part Number 1460005 Revision C ...