SEC 3120 16-bit Modbus Manual
Page 21 Of 46
Sensor Electronics Corporation
For Version 1.2 and above
December 2012
P/N 1580282, Rev. 1, 20121228
Holding
Register
Address
Read or
Write
Sensor Register Description
(See Parameter Table for specifics)
42x28 -x 30
Read/Write
Calibration Value (IEEE 754 32-bit float)
42x31 -x32
Read
Sensor Range (IEEE 754 32-bit float)
42x33
Read
Alarm Configuration (bit map)
42x34 - x37
Read
Gas Name (ASCII characters eight, 1st left to right)
42x38 - x39
Read
Gas Units Name (ASCII characters four, 1st left to right)
42x40
Write Only
Start Self Test (with password)
Table 4: Region 9: 3120 Unit Registers
Holding
Register
Address
Read or
Write
3120 Unit Register Description
(See Parameter Table for specifics)
42901
Read
Unit Operating Status (bit map)
42902 - 03
Read
Fault Relay Reason Code (32-bit bit map)
42904 - 05
Read
Unit Serial Number (32-bit integer)
42906 - 11
Read
Clock Read Registers (real-time)
42912 - 17
Write Only
Clock Write Registers (New write hold registers)
42918
Write Only
Set Clock from New Clock write hold registers
42919
Write Only
Diagnostics (force alarm/fault relays, force LED blinking)
42920
Write Only
Set Unit into Listen Only Modbus Mode
42921
Write Only
Restore Unit back from Listen Only Modbus Mode
42922
Read
Maximum number of sensors that can be attached
42923
Read
Unit Type Code
42924 - 25
Read
Unit Firmware Version (2- words for major.minor.build)
Modbus Protocol
Introducing Modbus Protocol
Modbus communication is based on a master–slave technique, in which only one
device (the master) can initiate transactions (queries). The other devices (the slaves)
respond by supplying the requested data to the master, or by taking the action requested
in the query. The master can address individual slaves or can initiate a broadcast message
to all slaves. Slaves return a message (a response) to queries that are addressed to them
individually. Responses are not returned to broadcast queries from the master.
The Modbus protocol establishes the format for the master’s query by placing the
device (or broadcast) address, a function code defining the requested action, any data to
be sent and an error-checking field into the message. The slave’s response message is
also constructed using Modbus protocol. The response contains fields confirming the
action taken, any data to be returned and an error-checking field. If an error occurred in
receipt of the message, or if the slave is unable to perform the requested action, the slave
will construct an error message and send it as the response.
Summary of Contents for SEC 3000
Page 15: ...Doc 1460003 Rev A Operation Manual SEC 3000 Electro Chemical Gas Sensor Page 12 of 13 ...
Page 54: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 4 SEC 3100 SEC Millenium Mounting ...
Page 55: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 5 SEC 3100 SEC 3000 Mounting ...
Page 56: ...TB 1 J1 TB 2 J2 1 6 1 6 1 4 1 6 TB 5 J5 TB 6 J6 ...
Page 148: ...SEC 3120 Wiring TB 1 P3 TB 2 P6 1 6 1 6 1 2 3 4 1 2 3 4 7 6 5 4 3 2 1 TB 3 P1 TB 4 P5 ...
Page 156: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 399: ...SEC P N 1460014 Rev ECO A 000209 Page 11 of 26 ...
Page 400: ...SEC P N 1460014 Rev ECO A 000209 Page 12 of 26 ...
Page 431: ...Part Number 1460005 Revision C ...
Page 433: ...Part Number 1460005 Revision C ...
Page 434: ...Part Number 1460005 Revision C ...
Page 435: ...Part Number 1460005 Revision C ...