SEC 3120 16-bit Modbus Manual
Page 39 Of 46
Sensor Electronics Corporation
For Version 1.2 and above
December 2012
P/N 1580282, Rev. 1, 20121228
Internal
Register
Address
Holding
Register
Address
Function Code 16- Write Multiple Sensor Registers
(Data bits and Parameter Specifics)
x = sensor number, or
0 to apply to all sensors.
*The real value must be multiplied by the Gas Factor (2x08).
i.e. 2.12 ppm to set low set point register: scaled low set point = 2.12 X 10 (from 2x08) =
212 (00D4 Hex). When the password is added (4000 Hex) the complete value needed to
write the register and change the low set point then is 40D4 Hex.
**Sensor Float parameters (low, mid and high set points and calibration values) must be
written in sequence in at least one multiple item write operation, starting in order with the
enable password register, high word then low word. These values are IEEE 754 encoded
32-bit values split into two 16-bit words. Upon a complete write of the triad, the value is
saved internally.
***Zone ID (2010/42011) and Self Test (2039/42040) may be written in region 0 (apply
to all sensors) when a SEC 3120 is in either "Identical" or "Unique" sensor roles (unlike
set point or threshold registers where the sensor role must be "Identical"), however if in
the "Single" sensor role, an exception message will be transmitted to the master. If sensor
role is "Identical", values must also be identical!
Internal
Register
Address
Holding
Register
Address
Function Code 16- Write Multiple Unit Registers
(Data bits and Parameter Specifics)
2911
42912
New 3120 Unit Clock Month Value.
4000 Hex must be added as a password to write value.
2912
42913
New 3120 Unit Clock Day Value.
4000 Hex must be added as a password to write value.
2913
42914
New 3120 Unit Clock Year Value.
4000 Hex must be added as a password to write value.
2914
42915
New 3120 Unit Clock Hours Value.
4000 Hex must be added as a password to write value.
2915
42916
New 3120 Unit Clock Minutes Value.
4000 Hex must be added as a password to write value.
2916
42917
New 3120 Unit Clock Seconds Value.
4000 Hex must be added as a password to write value.
2917
42918
Set New 3120 Unit Clock NOW (00FF Hex set, 0 clear).
4000 Hex must be added as a password to write value.
2918
42919
Bit-mapped Diagnostics Force Coils And Toggle LEDs.
Bit 0: Force low relay coil (1 = on, 0 = off)
Bit 1: Force mid relay coil (1 = on, 0 = off)
Bit 2: Force high relay coil (1 = on, 0 = off)
Bit 3: Force fault relay coil (1 = on, 0 = off)
Bit 4: Toggle LEDs red/green ( 1 = on, 0 = off)
Duration of force will be approximately 30 seconds unless
re-written.
4000 Hex must be added as a password to write value.
Summary of Contents for SEC 3000
Page 15: ...Doc 1460003 Rev A Operation Manual SEC 3000 Electro Chemical Gas Sensor Page 12 of 13 ...
Page 54: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 4 SEC 3100 SEC Millenium Mounting ...
Page 55: ...Part Number 3100 EXP MANUAL REV 11 20141007 FIGURE 5 SEC 3100 SEC 3000 Mounting ...
Page 56: ...TB 1 J1 TB 2 J2 1 6 1 6 1 4 1 6 TB 5 J5 TB 6 J6 ...
Page 148: ...SEC 3120 Wiring TB 1 P3 TB 2 P6 1 6 1 6 1 2 3 4 1 2 3 4 7 6 5 4 3 2 1 TB 3 P1 TB 4 P5 ...
Page 156: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 399: ...SEC P N 1460014 Rev ECO A 000209 Page 11 of 26 ...
Page 400: ...SEC P N 1460014 Rev ECO A 000209 Page 12 of 26 ...
Page 431: ...Part Number 1460005 Revision C ...
Page 433: ...Part Number 1460005 Revision C ...
Page 434: ...Part Number 1460005 Revision C ...
Page 435: ...Part Number 1460005 Revision C ...