85
10. Press the “Y Mask” button; turn the “Universal” knob to adjust the vertical clearance.
11. Press the “Create Mask” button to create mask and you also can enter the next page of “Mask menu”
to recall the stored mask.
12. Enter the second page of “Pass/Fail function menu”, press the “Output” option button to set the
output option.
13. Enter the first page of “Pass/Fail function menu”, press the “Operate” option button to select “ ”to run
the pass/fail test.
h) Waveform Record
Waveform record can record input waveform from CH1 and CH2 with a maximum record length of 2500
frames. This record behavior can also be activated by the pass/fail test output, which makes this function
especially useful to capture abnormal signals in long term without keeping an eye watching it. Waveform
recorder: Record the waveforms with specified interval.
Fig. 73 Pass Fail Setup Page 1
Table 61: Waveform recorder function menu
User Manual SMO702, SMO1002