47
f) Trigger Holdoff
You can use the Trigger Holdoff function to produce a stable display of complex waveforms. Holdoff is time
between when the oscilloscope detects one trigger and when it is ready to detect another. The oscilloscope
will not trigger during the holdoff time. For a pulse train, you can adjust the holdoff time so the oscilloscope
triggers only on the first pulse in the train.
If you want to change holdoff time, please follow next operations:
1. Press the “TRIG MENU” button to show the “TRIG Menu”.
2. Press the “Type” option button to select trigger type.
3. Press the “Set up” option button to enter the “Trigger setup menu”.
4. Press the “Holdoff” option button, turn the “Universal” knob to change the holdoff time until the
waveform trigger steadily.
Note: Use trigger holdoff to help stabilize the display of aperiodic waveforms.
Fig. 33 Trigger Holdoff
User Manual SMO702, SMO1002