31
d) Using Ref
The reference control saves waveforms to a nonvolatile waveform memory. The reference function
becomes available after a waveform has been saved.
Press the Ref button to display the “Reference waveform menu”.
Operation step :
1. Press the “REF” menu button to display the “Reference waveform menu”.
2. Press the “Source” option button to select input signal channel.
3. Turn the vertical “POSITION” knob and “Volt/div” knob to adjust the vertical position and scale to
conformable positions.
4. Press the third option button to select “REFA” or “REFB” as storage position.
5. Press the “Save” option button.
6. Press the bottom option button to select “REFA On” or “REFB On” to recall the reference waveform.
Table 11: REF Function Menu
Fig.
21
Reference Waveform Menu
User Manual SMO702, SMO1002