Scientific
SMO1000E
53/185
Select Waveform Interpolation Method
Under real-time sampling, the oscilloscope acquires the discrete sample values of the
waveform being displayed. In general, a waveform of dots display type is very difficult to
observe. In order to increase the visibility of the signal, the digital oscilloscope usually
uses the interpolation method to display a waveform.
Interpolation method is a processing method to “connect all the sampling points”, and
using some points to calculate the whole appearance of the waveform. For real-time
sampling interpolation method is used, even if the oscilloscope in a single captures only a
small number of sampling points. The oscilloscope can use interpolation method for filling
out the gaps between points, to reconstruct an accurate waveform.
Press the
Acquire
button on the front panel to enter the ACQUIRE Function menu; then
press the
Interpolation
softkey to select
Sinx/x
or
X
.
X:
In the adjacent sample points are directly connected on a straight line. This
method is only confined to rebuild on the edge of signals, such as square wave.
Sinx/x:
Connecting the sampling points with curves has stronger versatility. Sinx
interpolation method uses mathematical processing to calculation results in the actual
sample interval. This method bending signal waveform, and make it produce more
realistic regular shape than pure square wave and pulse. When the sampling rate is 3 to
5 times the bandwidth of the system. Recommended Sinx/s interpolation method.
Figure 11 Display Type Set to Dots
Summary of Contents for SMO1000E
Page 1: ...Digital oscilloscope SMO1000E User Manual ...
Page 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Page 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Page 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Page 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Page 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...