Figure 36
7 Address & Data
— the oscilloscope will be triggered when the following conditions
are satisfied.
The address’s length must be 7 bits and the address’s value is the same as set
value.
If you have set either Data1’s or Data2’s value, and the signal has a data is the same
as that value. If you have set both Data1’s and Data2’s value, the signal should has
two consecutive data, the first data’s value is Data1, second data value is Data2.
(
Note
: If the data’s value is 0xXX, any data value will be matched)
Figure 37 On the 17th Clock Edge
Figure 38 On the 26th Clock Edge