![Scientific SMO1000E User Manual Download Page 112](http://html.mh-extra.com/html/scientific/smo1000e/smo1000e_user-manual_1217342112.webp)
Scientific
SMO1000E
112/185
CAN Trigger and Serial Decode
Please in order of “
Setup for CAN Signals
”, “
CAN Triggering
” and “
CAN Serial
Decode
” to trigger and decode the signals.
Setup for CAN Signals
1. Press the
Decode
key to enter the
DECODE
function menu.
2. Press the
Decode
softkey and select the desired slot (Decode1 or Decode2).
3. Press
Protocol
softkey and then select
CAN
by turning Universal Knob.
4. Press
Signal
softkey to enter the
SIGNAL
menu as Figure 55 shows.
Figure 55 CAN SIGNAL Menu
5. Set CAN-H:
a. Press
CAN-H
softkey to select the channel that is connected to the CAN-H
signal.
b. Press first
Threshold
key to set the CAN-H signal’s threshold voltage level by
Universal Knob
. The threshold voltage level is for decoding, and it will be
regard
as the trigger voltage level when set the trigger type to serial.
6. Set CAN-L:
a. Press
CAN-L
to select the channel that is connected to the CAN-L signal.
b. Press second
Threshold
key to set the CAN-L signal’s threshold voltage level
by
Universal Knob
. The threshold voltage level is for decoding, and it will be
regard
as the trigger voltage level when set the trigger type to serial.
7. Press the
Configure
softkey to enter the
BUS CONFIG
menu.
8. Press
Baud
to set baud rate by Universal Knob.
The baud rate can be set as predefined value (from 5kb/s to 1Mb/s) or custom value
(from 1b/s to 1Mb/s).
If the desired baud rate is not listed, press
Baud
and select
custom
option, press
the
Custom
and turn the Universal Knob to set the desired baud rate.
Summary of Contents for SMO1000E
Page 1: ...Digital oscilloscope SMO1000E User Manual ...
Page 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Page 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Page 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Page 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Page 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...