SMO100E
97/188
Figure 40 I2C TRIGGER Menu
5. Press the
Condition
softkey and turn the Universal Knob to select the trigger
condition:
If you select the
EEPROM
condition:
a. Press the
Limit Range
softkey to set the qualifier (
=
,
<
or
>
).
b. Press
Data1
softkey and set its value by turning the Universal Knob.
If you select
7 Addr & Data
or
10 Addr & Data
condition:
a. Press the
Addr
softkey and turn the Universal Knob to select the 7- bit or
10-bit device address.
b. Press the
Data1
or
Data2
softkey and set the value about them.
c. Press The
R/W bit
softkey and select write-frame or read-frame to trigger
the oscilloscope.
(
Tips
: If device address is 7-bit, the value of address is in range of 0x00 to 0x7F.If
device’s address is 10-bit, the value of address is in range of 0x00 to 0x3FF.)
If you select the
Data Length
condition:
a. Press
Address
to set the SDA address length 7bit or 10 bit.
b. Press
Byte Length
softkey and set the byte length by Universal Knob. The
range of the Byte Length is 1 to 12.