CAN Triggering
This part will provide a brief introduction and description for the operation of the LIN
trigger.
To introduce the trigger conditions
Start
— the oscilloscope will be triggered at the start bit of a frame.
Remote
— the oscilloscope will be triggered by a remote frame with specified ID.
ID
— the oscilloscope will be triggered by a remote or data frame that have specified ID.
ID+DATA
— the oscilloscope will be triggered by data frame that have specified ID and
data.
Error
— the oscilloscope will be triggered by an error frame.
Operation steps
:
1. Press
Setup
to enter the
TRIGGER
function menu.
2. Press
Type
and select
Serial
.
3. Press
Protocol
and select
CAN
.
4. Press
Trigger Setting
to enter the
CAN TRIG SET
menu.
5. Press
Condition
and select the trigger condition by Universal Knob:
If you select the
REMOTE
and
ID
condition:
a. Press
ID Bits
to set the length of ID (11bits or 29 bits).
b. Press
Curr ID Byte
and use Universal Knob to select the byte that you want to set.
c. Press the
ID
and set the ID’s value by Universal Knob.
(
Tips
: In order to make is convenient for the operator to set the parameters, ID b
is split into several bytes. For example, if the ID’s length is 11 bits, it will be split
into two bytes, a byte includes 8 bits. If “1st byte” is selected, only the 8 least
significant bits can be changed.)
Summary of Contents for SMO1000E
Page 1: ...Digital oscilloscope SMO1000E User Manual ...
Page 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Page 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Page 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Page 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Page 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...