To Do Pass/Fail Test
One way to verify a waveform's compliance to a particular set of parameters is to use
pass/fail testing. A pass/fail defines a region of the oscilloscope's display in which the
waveform must remain in order to comply with chosen parameters. Compliance to
the mask is verified point- by- point across the display.
The test results can be displayed on the screen as well as be declared through the
system sound or the pulse signal output from the
[TRIG OUT]
connector at the rear
panel. Pass/Fail test operates on displayed analog channels; it does not operate on
channels that are not displayed.
Figure 66 Pass/Fail Test