To Do Self Calibration
The self-calibration program can quickly make the oscilloscope reach the best working state
to get the most precise measurement values. You can perform self-calibration at any time
especially when the change of the environment temperature is up to or more than
5
℃
. Make sure that the oscilloscope has been warmed up or operated for more than
30 minutes before the self-calibration.
Do the following steps to do self calibration:
1. Disconnect all the input channels.
2. Press the
Utility
button on the front panel
,
and then press the
Do Self Cal
softkey,
and the oscilloscope will pop-out the message box shown as below:
Figure 65 Do Self Cal
3. Press the
Single
button on the front panel to perform the self calibration program.
During the calibration, most of the keys are disabled.
4. When the self calibration program is finished, it will pop-out the message
“
press
Run/Stop key to exi
t”. Press the
Run/Stop
button on the front panel to
exit the
calibration interface.
Summary of Contents for SMO1000E
Page 1: ...Digital oscilloscope SMO1000E User Manual ...
Page 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Page 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Page 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Page 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Page 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...