Scientific
SMO1000E
172/185
To Update Firmware and Configuration
The firmware and configuration can be updated directly via USB flash driver.
Do the following steps to update the firmware:
1. Insert the USB flash driver which with the firmware and the configure files inside into
the USB host interface on the front panel of the oscilloscope.
2. Press the
Utility
button on the front panel to enter the UTILITY function menu.
3. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
4. Press the
Next Page
softkey to go to the third page of the UTILITY function menu.
5. Press the
Update
softkey to enter the UPDATE function menu.
6. Press the
Firmware
softkey and it will pop-out the message “
Press
„Single
‟
to
continue and press „Run/Stop
‟
to exit!
”
7. Press the
Single
button to enter the SAVE/RECALL file system.
8. Turn the
Universal Knob
to select the update file which should be with an ADS
postfix; and then press the
Load
softkey to start update the firmware. The process
needs about 7 minutes. And during the update, do not cut off the oscilloscope,
otherwise the oscilloscope will not restart again.
9. After finish the update, the screen will pop-out the message “
Update success,
please
restart the DSO
”, and you will heard a sound of buzzer.
10. Restart the oscilloscope to finish the firmware update.
Do the following steps to update the configuration:
1. Insert the USB flash driver which with the firmware and the configure files inside
into the USB host interface on the front panel of the oscilloscope.
2. Press the
Utility
button on the front panel to enter the UTILITY function menu.
3. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
4. Press the
Next Page
softkey to go to the third page of the UTILITY function menu.
5. Press the
Update
softkey to enter the UPDATE function menu.
6. Press the
Configure
softkey to enter the SAVE/RECALL file system.
7. Turn the
Universal Knob
to select the update file which should be with an CFG
postfix; and then press the
Load
softkey to start update the firmware. The
process needs about 30 seconds.
8. After finish the update, the screen will pop-out the message “
Update success,
please
restart the DSO
”, and you will heard a sound of buzzer.
9. Restart the oscilloscope to finish the configuration update.
Summary of Contents for SMO1000E
Page 1: ...Digital oscilloscope SMO1000E User Manual ...
Page 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Page 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Page 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Page 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Page 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...