To Save and Recall Test Mask
Users can save the current test mask to the internal Flash memory or external USB
flash device. The file format of the test mask file is “*.RGU”.
Save Test Mask to Internal Memory
1. Press the
Utility
button on the front panel to enter the UTILITY function menu.
2. Press the
Next Page
softkey to go to the second page of the UTILITY
function menu.
3. Press the
Pass/Fail
softkey to enter the PASS/FAIL function menu.
4. Press the
Enable Test
softkey to select
On
to enable the pass/ fail test.
5. Press the
Source
softkey to select the desired channel.
6. Press the
Mask Setting
softkey to enter the MASK function menu.
7. Press the
X Mask
or
Y Mask
softkey; and then turn the
Universal Knob to
select the desired value.
8. Press the
Location
softkey to select
Internal
9. Press the
Save
softkey to save the mask to internal memory.
10. A few second later, it will pop-out the message “
Store Data Success!
”
Note: the internal memory can only save one test mask; save a new test mask
will cover the old one.
Recall Test Mask from Internal Memory
1. Press the
Utility
button on the front panel to enter the UTILITY function menu.
2. Press the
Next Page
softkey to go to the second page of the UTILITY
function menu.
3. Press the
Pass/Fail
softkey to enter the PASS/FAIL function menu.
4. Press the
Enable Test
softkey to select
On
to enable the pass/ fail test.
5. Press the
Mask Setting
softkey to enter the MASK function menu.
6. Press the
Location
softkey to select
Internal.
7. Press the
Load
softkey to recall the saved interval memory.
8. A few second later, it will pop-out the message “
Read Data Success!
” and
the saved mask will be displayed on the screen.
Summary of Contents for SMO1000E
Page 1: ...Digital oscilloscope SMO1000E User Manual ...
Page 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Page 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Page 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Page 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Page 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...