Instructions for use arco-matic 200 M / 300 M - Version 2018-05-08 GBR ID no. 2038823A
Feature
Value
Expected service life in accordance with IEC 60601-
1:2005 + amendment :2006 + amendment:2007 +
supplement1:2012, Section 3.28 (see “Glossary” on
page 82)
10 years, subject to compliance with the recom-
mended biannual inspections.
Safe working load as per IEC 60601-1:2005 +
amendment :2006 + amendment:2007 + supple-
ment1:2012 (maximum permitted patient weight
less the weight of the accessories attached to the
examination and treatment chair) (see“Glossary” on
page 82)
250 kg
Weight (without accessories)
96 kg ( arco-matic 200 M),
100 kg ( arco-matic 300 M ),
Degree of protection as per IEC 60529:1989 +
A1:1999 + A2:2013 (see “Glossary” on page 82)
IPX4 with closed safety cover,
IPX0 with open safety cover
Protection class as per IEC 61140:2001 (see “Glos-
sary” on page 82)
I
Suitability for use in oxygenated atmosphere
The examination chair must not be operated in an
oxygenated atmosphere.
Mains voltage
100-240 V AC
Frequency
50 Hz / 60 Hz
Power consumption (including sockets)
maximum 1.7 kW
Power output of the sockets
maximum 1.5 kW
Nominal duty type as per IEC 60034-1:2017 (see
“Glossary” on page 82)
Intermittent periodic duty (duty type S3)
INT 2 mins/18 mins (the device is not designed for
continuous duty. After 2 mins of full load opera-
tion, an 18 min break is required.)
Tab. 17
Technical data of the examination and treatment chair
Feature
Value
Short description
T6,3 A H 250 V
Characteristics
slow-blow
Breaking capacity
1,500 A
Measurements
5 × 20 mm
Tab. 18
Specifi cation for fi ne-wire fuses