24
Instructions for use arco-matic 200 M / 300 M - Version 2018-05-08 GBR ID no. 2038823A
Fitting the front foot section cover
Lift the panel as shown in Fig. 26.
Slide the front foot section cladding towards the
rear foot section cladding as shown in Fig. 26.
Fig. 26
Mounting the front foot section cladding
1) Front foot section cladding 2) Panel
Ensure that the panel does not rest at any point on
the collar of the foot section cover.
NOTE
If the panel rests on the collar of the front foot
section cladding, the cladding of the examination
and treatment chair will be damaged when the seat
pad is lowered.
Hence, ensure that the panel can slide without
obstruction over the collar of the front foot section
cladding!
Fig. 27
Faulty installation of the front foot section cladding
1) Panel 2) Collar of the front foot section cladding 3)
Point where the panel rests on the collar
Attach the front foot section cladding to the back
foot section cladding using the top and bottom
attachment screw as shown in Fig. 28.
Fig. 28
Attaching the front and rear foot section cladding
1) Top attachment screw 2) Bottom attachment screw