36
Instructions for use arco-matic 200 M / 300 M - Version 2018-05-08 GBR ID no. 2038823A
Displays
The display of the hand control unit has
•
two general warning lights,
•
one motor indicator light for every adjustable
element,
•
four memory lights and
•
the operational status light.
For the meaning of the light modes of the warning
lights, display lights and memory lights, see Tab. 9.
Lights
Colour
Symbol
Action to be taken
when LED is perma-
nently on
Action to be taken when LED is
fl ashing
General warning
light
yellow
(not applicable)
Excessive strain on motor
Reduce load.
General warning
light
yellow
(not applicable)
Controller reset: press and hold
buttons 1 and 2 on the hand control
unit together for approx. 5 seconds.
If the warning light continues
to fl ash amber: notify Technical
Service .
Malfunction indi-
cator lights
yellow
Notify Technical
Service.
Lower the seat section as far as
possible.
Lower the back section as far as
possible.
Lower the foot support as far as
possible.
The LED illumination stops blinking.
Memory lights
green
1 / 2 /
3 / 4
The memory position
was saved successfully.
The current position of the examination
and treatment chair will be saved as
the new memory position (press the
memory button).
Operational status
light
green
The examination and
treatment chair is
switched on.
(not applicable)
Tab. 9