Instructions for use arco-matic 200 M / 300 M - Version 2018-05-08 GBR ID no. 2038823A
Mobility (101.0466.0 )
“Mobility” refers to fi tting out the examination and
treatment chair with two double castors and four
fi xed-type castors as well as a pedal for central
locking. These fi ttings makes the examination and
treatment chair mobile.
Due to the fi xed-type castors and the double castors,
the height of the seat increases by 11 mm .
The examination and treatment chair cannot be fi tted
with mobility equipment if it is hardwired.
Safe working load: see table in chapter “Approved
accessories” on page 7.
DANGER!
Patients may incur severe injuries through collisions
or falling down when they are transported on the
examination and treatment chair.
Therefore, never move the examination and treat-
ment chair when patients are sitting on it!
DANGER!
If the mains plug is not disconnected, the power
cord may suddenly snap when moving the examina-
tion and treatment chair. People may be exposed to
a life-threatening electric shock as a result.
Disconnect the mains plug before moving the exami-
nation and treatment chair!
WARNING!
Unsecured equipment may move unexpectedly when
the patient is getting on, which can lead to injuries or
damage to property.
Hence, before the patient gets on, always ensure
that all castors on the device have been locked!
NOTE
The power cord and supply cable for the foot control
unit may be damaged if the double castors roll over
them.
Therefore, always ensure that the double castors
never roll over an electric cord.
Fig. 75
Pedal
1) Pedal