Instructions for use arco-matic 200 M / 300 M - Version 2018-05-08 GBR ID no. 2038823A
Operation
Safety information
WARNING!
When adjusting the device and accessories, body
parts may be crushed/overstretched.
For this reason, please observe the patient during
adjustment and ensure that such injuries are
avoided!
WARNING!
There is an increased risk of injury for patients, users
and third parties when using the memory function,
as it is diffi
cult to assess automatic adjustment in
relation to accessories and other objects. This can
lead to e.g. crushing injuries.
For this reason, keep the examination chair under
constant observation during automatic adjustment!
Make sure that you are able to step in to help at any
time! In order to cancel an adjustment, press any
button on the hand control or foot control.
WARNING!
During adjustment, the examination chair or the
accessories mounted on it (e.g. Göpel leg support)
may collide with body parts or other objects, leading
to personal injury or property damage.
Make sure that there is suffi
cient clearance when
performing adjustments for this reason!
CAUTION!
When the examination chair is lowered, the doctor’s
feet may be crushed if they are on the foot rest.
For this reason, the feet should be removed from the
foot rest before lowering the examination chair!
CAUTION!
Patients may slide off the seat surface if the Göpel
leg support is lowered all the way down.
For this reason, only lower the Göpel leg support ¾
of the way!
NOTE
While adjusting the device and accessories, objects
may be damaged.
Ensure suffi
cient clearance when performing adjust-
ments!
NOTE
When adjusting the examination and treatment chair,
mounted colposcopes can easily collide with the
chair or attached accessories and cause damage.
Before adjusting the examination and treatment
chair, swivel the mounted colposcope to the side!