68
Instructions for use arco-matic 200 M / 300 M - Version 2018-05-08 GBR ID no. 2038823A
Checklist
1. Has the product been cleaned and disinfected in accordance with the hygiene guidelines in eff ect at your
hospital/clinic?
Yes
No actions are required for this item.
No
Stop using the product for the time being.
Clean and disinfect the product.
2. Do the pads show signs of cracking or damage?
Yes
Stop using the pads.
Replace the pads.
No
No actions are required for this item.
3. Does the plastic cladding show damage?
Yes
Replace the plastic cladding.
No
No actions are required for this item.
4. Does the hand control unit or the foot control unit show any damage?
Yes
Notify Technical Service.
Replace the damaged control unit.
No
No actions are required for this item.
5. Does the mains feed cable or the on/off switch show any damage?
Yes
Notify Technical Service.
Professionally replace the damaged
components.
No
No actions are required for this item.
6. Is the examination and treatment chair wobbly? (Only if the examination and treatment chair is not
equipped with double castors.)
Yes
Adjust the levelling feet of the examina-
tion and treatment chair.
No
No actions are required for this item.
7. Operate the adjustment buttons of the hand/foot control unit one after the other. Do all adjustment func-
tions work properly?
Yes
No actions are required for this item.
No
Press adjustment buttons
and
on
the hand control unit at the same time
(reset).
8. Do all adjustment functions of the hand/foot control unit work properly after pressing the reset buttons?
Yes
No actions are required for this item.
No
Stop using the product for the time being.
Notify Technical Service.
9. Is the leg plate extractable?
Yes
No actions are required for this item.
No
Notify Technical Service.
Date
Signature of the inspector
Tab. 15
Checklist for visual and functional checks