Instructions for use arco-matic 200 M / 300 M - Version 2018-05-08 GBR ID no. 2038823A
27
Sockets on the device (optional, 101.0482.0 )
DANGER!
Risk of electric shock!
If a multi-socket outlet is installed, connecting elec-
trical devices to the multi-socket outlet constitutes
an ME system and thus a reduced level of safety
according to IEC 60601-1:2005+A1:2012. This has
consequences for the “Responsible Organisation”
(the operator).
The operator (doctor, clinic, hospital) must ensure
that the ME system (the examination and treatment
chair together with the connected devices) fulfi l
certain electrotechnical installation and operational
requirements in order to protect the patient and
medical personnel.
These requirements can be found in IEC 60601-
1:2005+A1:2012.
WARNING!
When connecting the mains plug, devices connected
to the integrated sockets can sometimes unex-
pectedly move and cause personal injury. This can
happen regardless of the position of the master
switch on the examination and treatment chair.
Therefore, before connecting the mains plug, make
sure that the devices connected to the integrated
sockets are safe to operate.
For example, examination lights and video colpo-
scopes can be connected to the integrated sockets
on the examination and treatment chair.
Voltage is applied to the integrated sockets on the
examination and treatment chair as soon as the
mains plug of the examination and treatment chair
is connected to the mains. If the examination and
treatment chair is hardwired to the mains, there is
always voltage at the integrated sockets.
Unlike other examination and treatment chairs
by SCHMITZ u. Söhne GmbH & Co.KG, it is not
possible to switch the integrated sockets on and off .
The sockets can be disconnected from the power
supply in the following ways:
•
Disconnect the mains plug (if the examination and
treatment chair is connected to the mains by a
mains plug),
•
Press the switch of the ground cable (if the
examination and treatment chair is hardwired to
the mains).
Connecting
Undo the attachment screws of the socket clad-
ding.
Remove the socket cladding.
Connect electric equipment, such as a colpo-
scope or examination light, to the integrated
socket.
Fig. 33
Remove the socket cladding
Fig. 34
Power supply on the device
1) Integrated socket