Colposcope support (101.0467.0)
WARNING!
The colposcope support is not designed to be used
as a support or lifting handle. When subjected to
a certain load, the colposcope support can come
off the examination chair. Serious injury may occur
when the examination chair falls as a result.
For this reason, do not use the colposcope support
as a support or lifting handle!
All common colposcopes can be attached with the
colposcope support.
Safe working load: see table in chapter “Approved
accessories” on page 7.
Mounting
NOTE
When adjusting the examination and treatment chair,
mounted colposcopes can easily collide with the
chair or attached accessories and cause damage.
Before adjusting the examination and treatment
chair, swivel the mounted colposcope to the side!
Remove the panel from the right or left side of the
foot cover.
Insert the colposcope support in the mount as
shown in Fig. 82.
Insert attachment screws and turn them tight.
Removal
Undo and remove the attachment screws.
Remove the colposcope support.
Remount the panel on the right or left side of the
foot cover.
Fitting a Leisegang Balance-o-Matic
Remove the bottom clamp (see Fig. 84).
Slide the two clamping pieces onto the
connecting tube.
Mount the Balance-o-Matic and clamping pieces
in the adapters on the colposcope arm.
Fasten the fastening fl ange with two attachment
screws and washers to the foot cover.
Mount the colposcope on the colposcope support
according to the manufacturer’s instructions.
Fig. 82
Colposcope support
1) Attachment screws
Fig. 83
Mounted colposcope
Fig. 84
Leisegang Balance-o-Matic
1) Clamping piece 2) Connecting tube
3) Clamping piece 4) Lower clamp