Operators Handbook DCU3 – System
5., Main Function „Calibration“
BAEDCU3 - Rev. 4.6 - 0504
5 - 6
5.5
Calibration of Turbidimetry Measurement
Calibration of the turbimetry measurement probe is a „single-step“ procedure evalulating the „zero de-
viation“ of the sensor is. For this the sensor is placed into particle- and bubble-free deinonized water.
The system calculates the turbidimetry as the arithmetical mean of the value measured during a de-
fined measuring time in terms of „absorption units, AU“. For calculation the system considers the zero
deviation and a selectable damping factor „DAMP”. To obtain stable turbidimetry measurements, the
damping factor „DAMP“ can be selected in 4 stages.
The operating display of the turbidimetry sensor includes the data field for absorption units (AU), the
percentage sensor raw signal [%] and the zero deviation against „0 AU“. This allows for an easy func-
tional check of the sensor. The lamp contained in the turbidimetry sensor can be switched on and off
via the „Sensor Calibration“ - menu, see LAMP-entry
1)
.
5.5.1 Operating
Display
Tag
Entry Displayed value or function, required input
TURB
AU
Display of process value in 0...6 AU (Absorbing Units)
ELEC
%
Display of percentage raw signal of the sensor
ZERO
%
Display of zero deviation after calibration
LAMP
1)
On/off Display and selection of switching state of lamp in the sensor
DAMP
1 ... 4 Display and input of signal damping. The input signal is proc-
essed as the mean value of a floating reference window.
Damping factors base upon 4 selectable reference windows:
- step 1 = 6 s; step 2 = 12 s; step 3 = 30 s; step 4 = 60 s
Start Calibration
Enabling the calibration routine
Additional touch keys, values and functions available in this display:
BUFFZ
AU
Display of zero solution
1)
Only Fundalux I – system, not applied with Fundalux II (available as option from 2004 on)
Summary of Contents for DCU 3
Page 2: ......
Page 4: ...Operators Handbook DCU3 System Introduction BAEDCU3 Rev 4 6 0504...
Page 8: ...Operators Handbook DCU3 System Contents BAEDCU3 Rev 4 6 0504...
Page 10: ...Operators Handbook DCU3 System 1 Design and Function BAEDCU3 Rev 4 6 0504 1 2...
Page 74: ...Operators Handbook DCU3 System 6 Main function Control Loops BAEDCU3 Rev 4 6 0504 6 40...
Page 80: ...Operators Handbook DCU3 System 7 Main function Batch Control BAEDCU3 Rev 4 6 0504 7 6...
Page 88: ...Operators Handbook DCU3 System 8 Main Function Recipes BAEDCU3 Rev 4 6 0504 8 8...
Page 106: ...Operators Handbook DCU3 System 9 Main function Maintenance BAEDCU3 Rev 4 6 0504 9 18...
Page 110: ...Operators Handbook DCU3 System 10 Main Function Documentation BAEDCU3 Rev 4 6 0504 10 4...
Page 114: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...