Operators Handbook DCU3 – System
14., Supplement
BAEDCU3 - Rev. 4.6 - 0504
14 - 5
Text of alarm line
Meaning
Remedial measures
„2.SEC/not inst./102“
1
st
serial interface extension MVME335
is missing or installed incorrectly
Check board address;
reset to delivery state;
send a claim not for FIB
„VDO32/not inst./102“
Digital output extension VDOT-32 is
missing or installed incorrectly
Check board address;
reset to delivery state;
send a claim not for FIB
„VDO32/DOXXX-YYY/ZZZ“ Error
of
the digital output channels
XXX-YYY or digital output extension
VDOT-32; ZZZ = error code
Check cable connection;
check supply voltages
„VDO32/DIXXX-YYY /XXX“ Error of the digital input channels XXX-
YYY or digital output extension VDOT-
32; ZZZ = error code
Check cable connection;
check supply voltages
„VDO64/not inst./102“
Digital output extension VDOT-64 is
missing or installed incorrectly
Check board address;
reset to delivery state;
send a claim not for FIB
„VDO64/DOXXXYYY/ZZZ“ Error
of
the digital output channels
XXX-YYY of digital output VDOT-64;
ZZZ = error code
Check cable connection;
check supply voltages
14.2
Trouble Shooting for the DCU-System
The Sartorius BBI Systems GmbH develops, manufactures and configures the DCU system with ut-
most care. If problems arise with the use of the DCU system (malfunctions, technical problems), the
user or the customer service should return a detailed description of the problem.
The operator or the customers´ service can use the form „Function bug handling“ which is available on
request, to describe the problem. If all aspects are considered this will enable the Sartorius BBI Sys-
tems GmbH to answer within the shortest possible time offering a solution for the problem.
If problems arise with the DCU system we recommend that the user or the customers´ service should
contact the responsible local service representative of Sartorius BBI Systems GmbH first. However, if
no local service representative is available, he can directly send the completed form to:
Sartorius BBI Systems GmbH
Dept. FA (Product engineering, automation projects)
Schwarzenberger Weg 73-79
D - 34212 Melsungen, Fed. Republic of Germany
phone +49 56 61 - 71 34 00, fax +49 56 61 - 71 37 02
14.3 Interlocking
Functions
Interlocking functions are finally configured and cannot be changed by the user. Outputs
controlled and locked by an interlocking function are indicated by the state message „lock“
in the entry for manual access of the corresponding menu.
The extend of interlocking functions is specific for the fermentor and will be recorded in the
configuration tables. These tables are included in the delivery of the DCU-system.
Summary of Contents for DCU 3
Page 2: ......
Page 4: ...Operators Handbook DCU3 System Introduction BAEDCU3 Rev 4 6 0504...
Page 8: ...Operators Handbook DCU3 System Contents BAEDCU3 Rev 4 6 0504...
Page 10: ...Operators Handbook DCU3 System 1 Design and Function BAEDCU3 Rev 4 6 0504 1 2...
Page 74: ...Operators Handbook DCU3 System 6 Main function Control Loops BAEDCU3 Rev 4 6 0504 6 40...
Page 80: ...Operators Handbook DCU3 System 7 Main function Batch Control BAEDCU3 Rev 4 6 0504 7 6...
Page 88: ...Operators Handbook DCU3 System 8 Main Function Recipes BAEDCU3 Rev 4 6 0504 8 8...
Page 106: ...Operators Handbook DCU3 System 9 Main function Maintenance BAEDCU3 Rev 4 6 0504 9 18...
Page 110: ...Operators Handbook DCU3 System 10 Main Function Documentation BAEDCU3 Rev 4 6 0504 10 4...
Page 114: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...