Operators Handbook DCU3 – System
5., Main Function „Calibration“
BAEDCU3 - Rev. 4.6 - 0504
5 - 5
5.4
Redox - Calibration
The Redox calibration is in fact a functional check. You will measure a calibration buffer and compare
the measured value with the reference indicated on the buffer bottle. If these values are different, the
Redox can be entered manually. The DCU - system then calculates a corresponding correction factor.
5.4.1 Operating
Display
Tag
Entry
Displayed value or function, required input
REDOX
mV
Display of combined electrode voltage
CHECK
mV
Display of combined electrode voltage of last calibration
Start Check Electrode
Enabling calibration of the Redox electrode
Additional touch keys, values and functions available in this display:
REFBU
mV
Enter reference voltage of redox buffer
5.4.2 Functional
Check
The electrode function is checked before it is mounted into the culture vessel and therefore prior to the
sterilization. Since the heat impact of sterilization or chemical reactions with compounds of the culture
medium may effect the measuring characteristics of the electrode its function should be checked regu-
larly, i.e. prior to any next use and, if wrong operation is suspected, after a fermentation run.
1.
Place Redox electrode in a beaker filled with the reference buffer
2.
Enable functional check via „Start check electrode“. Follow the instructions of the menu.
5.4.3 Special
Notes
If the measured buffer Redox and the value on the buffer bottle deviate by more than 6 mV
(about 3%) the electrode must be serviced. See documentation delivered with the electrode.
Summary of Contents for DCU 3
Page 2: ......
Page 4: ...Operators Handbook DCU3 System Introduction BAEDCU3 Rev 4 6 0504...
Page 8: ...Operators Handbook DCU3 System Contents BAEDCU3 Rev 4 6 0504...
Page 10: ...Operators Handbook DCU3 System 1 Design and Function BAEDCU3 Rev 4 6 0504 1 2...
Page 74: ...Operators Handbook DCU3 System 6 Main function Control Loops BAEDCU3 Rev 4 6 0504 6 40...
Page 80: ...Operators Handbook DCU3 System 7 Main function Batch Control BAEDCU3 Rev 4 6 0504 7 6...
Page 88: ...Operators Handbook DCU3 System 8 Main Function Recipes BAEDCU3 Rev 4 6 0504 8 8...
Page 106: ...Operators Handbook DCU3 System 9 Main function Maintenance BAEDCU3 Rev 4 6 0504 9 18...
Page 110: ...Operators Handbook DCU3 System 10 Main Function Documentation BAEDCU3 Rev 4 6 0504 10 4...
Page 114: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...