6., Main function „Control Loops“
BAEDCU3 - Rev. 4.6 - 0504
6 - 31
6.20
Gravimetric Dosing Controller
The „Flow controller“ is a precise gravimetric dosing controller. It will be used together with a balance
system. Balance systems suited and tested for the DCU - system are the MC1-type balance by Sarto-
rius, Fed. Rep. of Germany, or Mettler Balances type AM/PM/SM, offering a RS232-type interface.
The storage vessel will be placed on a balance system, which is connected to the DCU - system via
the serial interface. The flow controller controls a peristaltic pump (any commercially available pump
for continuous control) to supply substrate from the storage vessel into the culture vessel. The DCU -
system calculates the delivered volume considering the weight change. It can ensure a constant flow
by continuous control of the external pump.
This control function directly acts upon the input weight signal delivered from the balance system.
Therefore the gravimetric dosing controller provides a precise dosing for days and weeks.
6.20.1 Operating
Display
For a description of the tags, the physical units or entries, the displayed value or function
and of the required inputs see the table in section
6.3,
Controller Operation in General.
6.20.2 Operating
Information
To operate the substrate supply with the flow controller proceed as follows:
1.
Fill storage vessel, sterilize, if necessary, and place onto balance platform
2.
Adjust the zero point of the balance
3.
Adjust the setpoint for the flow controller in the DCU - system
4.
Switch operating mode of the flow controller to „auto“
Check display of the weight, a negative value indicates the delivered volume (weight)
Summary of Contents for DCU 3
Page 2: ......
Page 4: ...Operators Handbook DCU3 System Introduction BAEDCU3 Rev 4 6 0504...
Page 8: ...Operators Handbook DCU3 System Contents BAEDCU3 Rev 4 6 0504...
Page 10: ...Operators Handbook DCU3 System 1 Design and Function BAEDCU3 Rev 4 6 0504 1 2...
Page 74: ...Operators Handbook DCU3 System 6 Main function Control Loops BAEDCU3 Rev 4 6 0504 6 40...
Page 80: ...Operators Handbook DCU3 System 7 Main function Batch Control BAEDCU3 Rev 4 6 0504 7 6...
Page 88: ...Operators Handbook DCU3 System 8 Main Function Recipes BAEDCU3 Rev 4 6 0504 8 8...
Page 106: ...Operators Handbook DCU3 System 9 Main function Maintenance BAEDCU3 Rev 4 6 0504 9 18...
Page 110: ...Operators Handbook DCU3 System 10 Main Function Documentation BAEDCU3 Rev 4 6 0504 10 4...
Page 114: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...