Operators Handbook DCU3 – System
7., Main function „Batch Control“
BAEDCU3 - Rev. 4.6 - 0504
7 - 4
7.2.3
Operation Automatic Sterilization Control
1.
Carry out the operations at the fermentor required for preparing the sterilization sequence,
see the Operating Manual of the fermentor for details. If necessary, adjust the sterilization
parameters (sterilization temperature, time etc.)
2.
Start the sterilization program with [STATE: start]. When requested by the system, carry out
a necessary manual step according to instructions on the display and confirm the message.
3.
Confirm the message „Sterilization finished“ with the [ACK] key after completion of the
automatic sterilization process.
If necessary, you can stop the automatic sterilization with [STATE: stop]. This automatically
initiates the cooling phase.
7.2.4 Special
Notes
During the running sterilization program the top line displays the process status
„STATE:active“.
If the process time has not yet been activated, it will automatically start when the sterilization
program is started.
Running sterilization programs can be stopped at any time. In this case the cooling phase is
automatically started to cool down the fermentor to ambient or the previously adjusted proc-
ess temperature as soon as possible.
If required an interrupted sterilization program can be restarted before the fermentor
reaches the preset process temperature.
In some fermentors a separate temperature controller is configured for the control of the
sterilization temperature.
If the display shows „STATE: lock“, the sequence is locked and cannot be started. Another
program (sequence, recipe or batch control) is active at this time. The lock will be released
and the program starts, when the running program is finished. You will find the specific in-
formation about the programs, which will lock the sequence due to equal or superior priority,
in the configuration documentation.
Summary of Contents for DCU 3
Page 2: ......
Page 4: ...Operators Handbook DCU3 System Introduction BAEDCU3 Rev 4 6 0504...
Page 8: ...Operators Handbook DCU3 System Contents BAEDCU3 Rev 4 6 0504...
Page 10: ...Operators Handbook DCU3 System 1 Design and Function BAEDCU3 Rev 4 6 0504 1 2...
Page 74: ...Operators Handbook DCU3 System 6 Main function Control Loops BAEDCU3 Rev 4 6 0504 6 40...
Page 80: ...Operators Handbook DCU3 System 7 Main function Batch Control BAEDCU3 Rev 4 6 0504 7 6...
Page 88: ...Operators Handbook DCU3 System 8 Main Function Recipes BAEDCU3 Rev 4 6 0504 8 8...
Page 106: ...Operators Handbook DCU3 System 9 Main function Maintenance BAEDCU3 Rev 4 6 0504 9 18...
Page 110: ...Operators Handbook DCU3 System 10 Main Function Documentation BAEDCU3 Rev 4 6 0504 10 4...
Page 114: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...