Operators Handbook DCU3 – System
6., Main function „Control Loops“
BAEDCU3 - Rev. 4.6 - 0504
6 - 9
6.6
Stirrer Speed Controller
The speed of the stirrer motor is controlled via an external, extra motor controller. You will adjust all
stirrer speed parameters at the DCU3 system. The control function of the DCU3 system works as a
setpoint controller and supplies the analog setpoint signal to the motor controller. Furthermore the
menu displays the stirrer speed signal delivered from the motor controller.
When the stirrer speed controller is switched to „off“, an additional digital output switches off the motor
contactor. If the fermentor is equipped with a pO
2
- controller the stirrer speed controller can be
switched as a servo controller in the pO
2
cascade control loop.
6.6.1 Operating
Display
For a description of the tags, the physical units or entries, the displayed value or function
and of the required inputs see the table in section
6.3,
Controller Operation in General.
6.6.2 Special
Notes
In „manual“ mode, when entering the MIN/MAX output limits and entering data into OUT,
allways observe the permissible measuring range of the corresponding speed control:
– BIOSTAT
®
ED : 0 - 100 % = 0 - 1500 rpm
For further information on the permissible speed range of your fermentor check the preset
configuration of your DCU3 – system or refer to the configuration documentation of your fer-
mentor system delivered extra.
Basic stirrer speed and/or maximum permitted speed during operation as a servo controller
in pO
2
cascade control are entered via the MIN/MAX output limits.
Summary of Contents for DCU 3
Page 2: ......
Page 4: ...Operators Handbook DCU3 System Introduction BAEDCU3 Rev 4 6 0504...
Page 8: ...Operators Handbook DCU3 System Contents BAEDCU3 Rev 4 6 0504...
Page 10: ...Operators Handbook DCU3 System 1 Design and Function BAEDCU3 Rev 4 6 0504 1 2...
Page 74: ...Operators Handbook DCU3 System 6 Main function Control Loops BAEDCU3 Rev 4 6 0504 6 40...
Page 80: ...Operators Handbook DCU3 System 7 Main function Batch Control BAEDCU3 Rev 4 6 0504 7 6...
Page 88: ...Operators Handbook DCU3 System 8 Main Function Recipes BAEDCU3 Rev 4 6 0504 8 8...
Page 106: ...Operators Handbook DCU3 System 9 Main function Maintenance BAEDCU3 Rev 4 6 0504 9 18...
Page 110: ...Operators Handbook DCU3 System 10 Main Function Documentation BAEDCU3 Rev 4 6 0504 10 4...
Page 114: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...