![Sartorius CAIS1 Operating Instructions Manual Download Page 72](http://html1.mh-extra.com/html/sartorius/cais1/cais1_operating-instructions-manual_1199616072.webp)
Checkweighing
H
(Combics 2)
With this application, you can check whether the sample on the weighing platform
matches a target value or lies within a given tolerance range. Checkweighing also
makes it easy to fill sample materials to a specified target weight.
Selection and settings in the
Applic.
/ Applic.2/
CHECK.WG
menu.
Features
– Enter the nominal or target weight (set point) and the tolerance range delimiters
either using the keypad or by saving the weight value of a load on the platform.
– Enter the tolerance limits as absolute values (Min and Max), as a percentage
deviation from the target or as a relative deviation from the target.
Setting:
Applic./Applic.2/CHECK.WG/CHECK.RG
menu item 4.5.
– The target value can be taken over as a weighed value from a weighing
platform, and the upper and lower tolerance limits are defined as a percentage
deviation from the target value (setting 4.5.2). The following percentages can
be selected as the deviation: 0.1%, 0.2%, 0.5%, 1%, 1.5%, 2%, 3%, 5% or 10%,
selection using the
r
key.
– The target value, lower tolerance limit (minimum) and upper tolerance limit
(maximum) can applied as weighed values from the weighing platform (menu
item 4.5.1).
– The target value can be applied as weighed values and via asymmetrical percent
limits (menu item 4.5.3).
– The target value can be applied as weighed values and via relative weight limits
(menu item 4.5.4).
– Target and tolerance limits checked during input; values must conform to:
upper limit > target > lower limit > 1 digit.
– Checkweighing range: either 30% to 170% of the target, or from 10% to infinity.
– Results are shown on the main display, as a bar graph and LED as well as sent to
control output ports for further processing.
– Toggle the main display between “weight value" and “relation to the target" by
pressing the
w
key. For the limit value, if the weight in the readout is outside
the tolerance range, “LL" (too low) or “HH" (too high) is displayed
– Activate Info mode via the
I
key
– Automatic results printout (
Applic./Applic.2/CHECK.WG/CHECK.RG
menu
item 4.6).
– Automatic taring of container weight (
APPLIC./aut.tare
menu item 3.7).
– Automatic initialization when you switch on the scale with most recently saved
application data (
APPLIC./aut.Strt
menu item 3.8).
Combics 2 only
: You can assign different functions to the
c
key for deleting applications. When
you clear applications, you can delete either the data stored for all applications or
just selected data stored for the active application.
Setting:
Applic./CLER.CF
menu item 3.24
Tare function:
If you store a tare (weight value) by pressing the
)
key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting:
Applic. / TARE.FNC
menu item 3.25.1 (factory setting)
72
Operating Instructions Combics Indicators
Operation
Summary of Contents for CAIS1
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Page 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Page 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Page 158: ...158 Operating Instructions Combics Indicators...
Page 159: ...Operating Instructions Combics Indicators 159...