..., 50%, 100%), configured in Setup under:
Activity
menu item 3.19.
If the “Averaging” parameter is set to 2%, for example, and the animal or object
weighs 10 kg, measurement does not begin until the fluctuation in weight value
remains below 200 g during three consecutive measurements.
Display
A calculated average value is shown continuously on the main display. The
G
symbol
indicates the calculated value.
You can toggle between the results display and the current scale display by pressing
the
w
key.
Setting:
Applic./Applic.1/ANIM.WG/DIS.UNLD
menu item 3.21.
You can select “Display is static until unload threshold reached” to have the display
switch automatically to the weight readout when you unload the weighing platform
(i.e., when the load is less than half the minimum load). The result of the most recent
averaging operation is not saved.
If you select “Display is static until the
c
key is pressed," the calculated average
remains displayed even after the weighing platform is unloaded, until you press the
c
key to begin a new measurement.
Example
: The weight of one mouse should be measured.
Configuration: The “Animal Weighing" application is selected, and printout has been
set up (see “Configuration").
t
Place empty container on the scale.
)
t
Tare the scale.
This is not required if the automatic tare function is active. The tare weight is
saved automatically when you place the container on the platform.
t
Place the mouse in the container.
2 0
t
Enter the number of sub-weighing operations using the keypad (in this example,
20 measurements).
O
t
Save the value entered and start the averaging.
Operating Instructions Combics Indicators
65
Operation
Summary of Contents for CAIS1
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Page 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Page 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Page 158: ...158 Operating Instructions Combics Indicators...
Page 159: ...Operating Instructions Combics Indicators 159...