![Sartorius CAIS1 Operating Instructions Manual Download Page 42](http://html1.mh-extra.com/html/sartorius/cais1/cais1_operating-instructions-manual_1199616042.webp)
Adjustment/Configuration counter for standard scales
Purpose
Automatic recording of changes to adjustment and weighing parameters using two
independent counters. The values remain saved for the life of the component.
t
To display both counters, press and hold the
(
key for longer than 2 seconds.
y
The “Configuration counter" is then shown in the weight display for 3 seconds
(identified by a
P
). Then the “Adjustment counter" is displayed for another
3 seconds (identified by a
C
). After 6 seconds, the information display turns off
automatically.
Adjustment counter features:
–
Counter limited to 9999
–
Counter at “C 0000" for hardware commissioning
–
Counter cannot be reset
–
Counter is updated automatically when:
– linearization, calibration/adjustment is successful
– user calibration, adjustment or linearization weight is changed (menu 1.18.)
– When the following parameters are changed:
Function of the
q
key (menu item 1.9.)
Zero setting range (menu item 1.11)
Tare/zero at power on (menu item 1.12)
The above parameters are reset to factory settings (menu item 9.1.1)
Configuration counter features:
–
Counter limited to 9999
–
Counter at “P 0000" for hardware commissioning
–
Counter cannot be reset
–
Counter is updated automatically when:
– When the following parameters are changed:
Installation location (menu item 1.1.)
Application filter (menu item 1.2.)
Stability range (menu item 1.3.)
Taring (menu item 1.5)
Auto zero (menu item 1.6.)
Weight unit 1 (menu item 1.7.)
Weight unit 2 (menu item 3.1.)
Weight unit 3 (menu item 3.3.)
The above parameters are reset to factory settings (menu item 9.1.1)
– Switching the
k
key to or from a 10-fold higher resolution
– Turning the application automatic taring on/off
(menu item 3.7.)
– The application parameters are reset to factory settings
(menu item 9.1.1)
42
Operating Instructions Combics Indicators
Operation
Summary of Contents for CAIS1
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Page 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Page 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Page 158: ...158 Operating Instructions Combics Indicators...
Page 159: ...Operating Instructions Combics Indicators 159...