Data Interfaces
Data Input Format
You can connect a computer to your scale to send commands controlling weighing
instrument functions and applications via the interface port.
All commands use the same data input format. They start with the
ESC
character
(ASCII 27) and end with a carriage return (
CR
; ASCII 13) and a line feed (
LF
; ASCII
10). The total length of a command is anywhere from 4 characters (1 command
character between the start and end described above) to max. 7 characters
(4 command characters). This number can also be higher when sending texts.
The commands listed in the following table must each be supplemented with ESC
...
CR LF.
Example
: The command character for output is “P” (“output to Port”). To trigger this
command, send the string: “ESC P CR LF”.
Command Meaning
K
Weighing mode 1
L
Weighing mode 2
M
Weighing mode 3
N
Weighing mode 4
O
Block keys
P
Send display value to data interface
Q
Output acoustic signal
R
Unblock keys
D
Tare and zero
(combination tare function)
f3_
Zero (see also the “kZE_” command)
f4_
Tare without zeroing (see also the “kT_” command)
i_
Information about the indicator, example of output: “C2/016202/1"
Explanation: Indicator: Combics 2, software version: 016202, active weighing platform: 1
kF1_
F1: Trigger
k
key function
kF2_
F2: Trigger
c
key function (Combics 2 only)
kF3_
F3: Trigger
r
key function (Combics 2 only)
kF4_
F4: Trigger
O
key function (Combics 2 only)
kF5_
F5: Trigger
w
key function (Combics 2 only)
kF6_
F6: Trigger
I
key function (Combics 2 only)
kF7_
F7: Trigger
g
key function (Combics 2 only)
kF8_
F8: Trigger
h
key function (Combics 2 only)
kF9_
F8: Trigger
R
key function (Combics 2 only)
kF10_
F8: Trigger
J
key function (Combics 2 only)
kF11_
F8: Trigger
K
key function (Combics 2 only)
kF12_
F8: Trigger
L
key function (Combics 2 only)
kCF_
CF: Trigger
c
key function (Combics 2 only)
kP_
Trigger
p
key function Print at printer interface
kT_
Trigger
T
key (tare)
kNW_
Trigger
n
key function (toggle the weighing platform)
kZE_
Trigger
(
key function (zero the instrument)
x1_
Output model designation of active weighing platform, example: “LP6200S-0C"
x2_
Output serial number of active weighing platform, example: “0012345678"
x3_
Output software version of active weighing platform, example: “00-42-04"
x4_
Output software version of indicator, example: “01-62-01"
x9_
Output serial number of indicator, example: “0012345678"
x10_
Output model of indicator, example: “CAW2P4-1500RR-LCE"
z1_
Input: printout header 1
z2_
Input: printout header 2
txx...x_
xx...x: text input, length corresponds to input, for display in the measured value line
The ASCII code for the “underline” character (“_”) is 95.
Format for entering printout header lines: “ESC z x a ... a _ CR LF" with x=1 or 2 and
a ... a: 1 to 20 characters for header x, followed by the underline, CR and LF
characters.
Summary of Contents for CAIS1
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Page 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Page 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Page 158: ...158 Operating Instructions Combics Indicators...
Page 159: ...Operating Instructions Combics Indicators 159...