![Sartorius CAIS1 Operating Instructions Manual Download Page 48](http://html1.mh-extra.com/html/sartorius/cais1/cais1_operating-instructions-manual_1199616048.webp)
Features
Which of the following features are available depends on the
connected weighing platform. These features are
configured in the
Setup
menu:
– External calibration/adjustment blocked in verified
weighing instruments
– External calibration/adjustment with the default weight
value or standard weight (not available on verified
instruments):
setup / wp-1
menu Menu item 1.9.
“Calibration, Adjustment."
– Specify the weight for external calibration/adjustment:
setup / wp-1
menu Menu item 1.18. “Entering
Calibration Weight."
– Internal adjustment for IS weighing platforms
(configure under:
COM1
or
unicom / WP2
), Combics 2
48
Operating Instructions Combics Indicators
Operation
(
1.) Zero the scale.
J
2.) Start calibration
(e.g., when adjustment
prompt flashes
WP
).
cal.Ext.
is displayed for
two seconds.
You are prompted to place
the required weight on the
platform (e.g., 10,000 g).
3.) Place the calibration/-
adjustment weight on the
weighing platform.
The difference between
the measured value and the
true weight of the sample
will be displayed with plus/
minus signs.
Ext. calibration
A printout will be
Nom. + 10000 g
generated if the process
Diff. + 1 g
is cancelled using the
(
J
4.) Activate calibration adjust-
ment (press the
(
key to
cancel).
The adjustment weight
is displayed once adjust-
ment is finished.
--------------------
A GMP-compliant
24.02.2012 10:15
printout is generated.
Type CAISL2
Ser.no. 12345678
Vers. 1.0103.11.2
BVers. 01-26-02
--------------------
Ext. calibration
Tar + 10000 g
Diff. + 1 g
Ext. adjustment
Diff. + 0 g
--------------------
24.02.2012 10:15
Name:
--------------------
– Block the
J
key to prevent use of the two functions
described above:
setup / wp-1
menu Menu item 1.9.
“Calibration, Adjustment."
– Calibrate first; then adjust automatically or manually (not
for verified weighing instruments):
setup / wp-1
menu
Menu item 1.10. “Calibration/Adjustment Sequence."
– Flashing
n
symbol as adjustment prompt. If more than
one weighing platform is connected, the platform number
is also displayed:
setup / wp-1
menu Menu item 1.15.
“Calibration Prompt."
– Block external calibration/adjustment:
setup / wp-1
menu
Menu item 1.16. “External Adjustment."
Example:
External calibration and manual adjustment with default weights (weighing parameters: factory settings)
Summary of Contents for CAIS1
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Page 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Page 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Page 158: ...158 Operating Instructions Combics Indicators...
Page 159: ...Operating Instructions Combics Indicators 159...