Available parameter settings
* = Factory setting
MIN.INIT
Minimum load for initialization
3.6
1 digit
1 digit*
3.6.1
2 digit
2 digits
3.6.2
5 digit
5 digits
3.6.3
10 digT.
10 digits
3.6.4
20 digT.
20 digits
3.6.5
50 digT.
50 digits
3.6.6
100 diG.
100 digits
3.6.7
200 diG.
200 digits
3.6.8
500 diG.
500 digits
3.6.9
1000 dI.
1000 digits
3.6.10
START
Start of averaging
3.18.
manual
Manual*
3.18.1
automat
Automatic
3.18.2
ACTIVTY
Animal activity
3.19.
0.1 perc.
0.1% of animal/object
3.19.1
0.2 perc.
0.2% of animal/object*
3.19.2
0.5 perc.
0.5% of animal/object
3.19.3
1 perc.
1% of animal/object
3.19.4
2 perc.
2% of animal/object
3.19.5
5 perc.
5% of animal/object
3.19.6
10 perc.
10% of animal/object
3.19.7
20 perc.
20% of animal/object
3.19.8
50 perc.
50% of animal/object
3.19.9
100 perc.
100% of animal/object
3.19.10
Autom. printout of results
3.20.
manual
Off*
3.20.1
automat
On
3.20.2
DIS.UNLD
Static display of result after load removed
3.21.
CLEARED
Display is fixed until unload
threshold reached*
3.21.1
PRESENT
Fixed display until
c
is pressed
3.21.2
)
t
To save the setting, press the
)
key.
( (
...
t
To exit setup: Press the
(
key several times.
Minimum load for initialization
You can set the minimum load here, i.e. the load that must be placed on the
weighing platform in order to carry out the application. If the load on the platform is
too light, the following will occur:
– Error code
Inf 29
appears,
– The weighing platform is not initialized,
– The preset reference sample quantity is saved.
Setting:
Applic./Applic.1/ANIM.WG/Min.init
menu item 3.6.
The minimum load can be set in 10 steps from 1 to 1000 digits (see available
parameters). The “digits" here refer to the scale intervals for the connected weighing
platform. If the interval of the connected platform is 1 g, for example, and 1000
digits are required, you must place at least 1000 g (= 1000 intervals =1000 digits)
on the weighing platform for initialization.
Starting the Measurements
The averaging routine does not begin until the fluctuation in weight value remains
below a defined threshold over three consecutive measurements. The tolerance limit
is defined as a percentage of the animal or object weight (for example, 0.1%, 0.2%,
64
Operating Instructions Combics Indicators
Operation
Summary of Contents for CAIS1
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Page 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Page 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Page 158: ...158 Operating Instructions Combics Indicators...
Page 159: ...Operating Instructions Combics Indicators 159...