Example
: 100% of a sample material should be weighed.
Confi guration: The “Weighing in Percent" application is selected, and printout has
been set up.
t
Place empty container on the scale.
)
t
Tare the scale.
This is not required if the automatic tare function is active. The tare weight is
saved automatically when you place the container on the platform.
t
Add reference material to the container in accordance with reference percentage
value (in this example, 85 g).
O
t
Start the calculation of the reference weight by pressing the
O
key.
y
The calculation is based on the active net weight value and the reference
percentage value entered
If the weight is too light, an error code is shown in the main display
Inf 29
.
If this is the case, set the minimum load to a smaller number of digits.
t
Add additional material until the reference percentage value has been reached
(in this example, 100 g)
70
Operating Instructions Combics Indicators
Operation
Summary of Contents for CAIS1
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators Declarations of Conformity...
Page 129: ...Operating Instructions Combics Indicators 129 Declarations of Conformity...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Plates and Markings...
Page 132: ...132 Operating Instructions Combics Indicators Plates and Markings...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 155: ...Operating Instructions Combics Indicators 155 Appendix Guide to verification...
Page 158: ...158 Operating Instructions Combics Indicators...
Page 159: ...Operating Instructions Combics Indicators 159...